the
ultimate in remote-controlled handling of babies

| |
Simply
place a UNIVERSAL OCTOPAQUE® unit on the four wheels
of the OCTOROLL®, and the examination is ready to
be conducted from the control bay. Immobilization is safe, efficient
and fast. The next baby may even be pre-immobilized outside
the room on another OCTOPAQUE®, and brought inside
for immediate exchange on the OCTOROLL®. Upper
gastrointestinal series, colon, voiding cystograms, etc. can
be done without tedious loss of time.
The radiologist remains protected
from radiation, yet just fingertip away from the patient, while
assistants supervise from a safe distance.
The unit is totally wireless
since it is battery operated and controlled by radio signals.
The battery charge is sufficient for an average day's work,
which may even continue while recharging. Stops are instantaneous
and a brake keeps the baby perfectly still during exposures.
Light and compact, the unit is
put away on a shelf under continuous charge for immediate use
at any time.
|
|
| |
ADAPTS
INSTANTLY TO ANY OCTOPAQUE® unit
and to ANY X-RAY APPARATUS
|
 
| |
OCTOROLL®
UM-02 SPECIFICATIONS*
Approved
by the FDA, but not by CE
MOTOR MODULE
MOTORS: two 13-commutator DC of the highest quality. Instant
starts and stops.
GEARS: two worm type. Provide strong internal brake when
not rotating..
FRAME: strong aluminum made with computer precision.
BATTERY: sealed lead acid, 12 volts (supplied), with
a life expectancy of over three years..
BATTERY CONTROLLER: 13.8 volts DC, with blinking red
LED at 50% discharge, indicating need for recharge, to prolong
battery life.
RADIO RECEIVER: RF frequencies with 256 safety codes.
Resist local incoming interference, which must nevertheless
be accepted.
SWITCHES: A) Main with LED.. B) Local rotation switch,
dead man type, clockwise and counterclockwise, overrides remote
control module.
MAXIMUM WEIGHT: 16 kg. Or up to 25 kg with optional wider
and stronger board.
DIMENSIONS: 37 L X 18 H X 16.5 W cm; 4 kg; feet component:
28 L X 10.5 H. X 10 W cm; 800 gm.
REMOTE CONTROL MODULE
BATTERY: 9 V. alkaline (supplied), good for over one year.
A buzzer in the motor module warns when battery is low (7 volts).
SWITCHES: two dead man switches, for clockwise and counterclockwise
rotation. Buttons are recessed to prevent accidental operation.
Simultaneous activation of both switches causes a buzz, but
no action or damages.
COMPLIANCE: with FCC rules Part 15. Possible outgoing
local interference must be accepted in a 235 m. radius; verification
must be made before operation and modification of security codes
done if necessary.
CHARGER
TYPE: wall plug, CSA and UL approved. Rated 800 ma.
INPUT:
110 V. AC. OUTPUT: 15 to 17 volts AC. When connected, allows
operation of the unit.
220
volts adapter is available as an option.
OUTPUT: 15 to 17 volts AC. When connected, allows operation
of the unit.
*Subject to change without
notice.
** EXAMINATIONS CAN ALWAYS BE
IMMEDIATELY CONDUCTED MANUALLY BY INSTALLING
THE IMMOBILIZED BABY DIRECTLY ON THE X-RAY TABLE
*** This device is not to be used
in a magnetic resonance environment.
|
|
WARNING
Must be used only by qualified
personnel, according to appropriate procedures, and under the
responsibility of a physician.
OCTOSTOP® Inc. and
its personnel do not assume any liability regarding the use,
indications, consequences, or any situation directly or indirectly
related to its products.
|
|